Factory Standard Calibration Information

Decision rule applied to all calibration certificates: MadgeTech Calibration Laboratory uses the shared risk approach as defined in ILAC G8 when making compliance statements.

Click here to see our current certificate and full scope of accreditation.

The following table provides the standard calibration points and the calibration certificate type provided for each MadgeTech data logger.

ProductStandard Calibration PointsCertificate Type
AVS140-1This product is a system, you must calibrate components separately. See HiTemp140N/A
AVS140-6This product is a system, you must calibrate components separately. See HiTemp140, PR140N/A
Bridge101A-1000mV
3 calibration points:
- Voltage at 0 mV
- Voltage at 1003 mV
- Voltage at -1003 mV
Standard Certificate
Bridge101A-150mV2 calibration points:
- Voltage at 0 mV
- Voltage at 100 mV
Standard Certificate
Bridge101A-30mV3 calibration points:
- Voltage at 0 mV
- Voltage at 24 mV
- Voltage at -24 mV
Standard Certificate
Bridge120-1000mV3 calibration points:
- Voltage at 0 V
- Voltage at 1 V
- Voltage at -498.0025 V
Standard Certificate
Bridge120-100mV3 calibration points:
- Voltage at 0 V
- Voltage at 0.1 V
- Voltage at -0.1 V
Standard Certificate
Bridge120-10mV3 calibration points:
- Voltage at 0 V
- Voltage at 0.01 V
- Voltage at -0.01 V
Standard Certificate
Bridge120-25mV3 calibration points:
- Voltage at 0 V
- Voltage at 0.025 V
- Voltage at -0.025 V
Standard Certificate
CryoTemp2 calibration points:
- Temperature at -80 °C
- Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
CTL2000
25 °C to 60 °C for Ambient channels, 0 mV for Thermocouple channels(No data included on certificate)Standard Certificate
DHS Data Logging System
This product is a system, you must calibrate it's components separately. See HiTemp140-M12N/A
EggTemp1 calibration point:
- Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
EggTemp-RH1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
Element CO21 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
1 calibration point:
- CO2 at 400 ppm (No data included on certificate)
Standard Certificate
Element HT1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
ETR101A2 calibration points:
- Temperature at 25 °C
- Temperature at 60.5 °C
1 calibration point:
- Thermocouple temperature at 0 mV
Standard Certificate
Event101AThe Event101A has a digital input signal and cannot be calibrated. A declaration of conformance can be provided.Declaration of Conformance
HiTemp140-12 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-22 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-2-TD2 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-5.252 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-TD-5.252 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-5.25-TD-TSK2 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-5.25-TSK2 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-72 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-7-TSK2 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-FP2 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-FR2 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-FR-TSK2 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-PT-12 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-PT-1-TSK2 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-PT-52 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-PT-5-TSK2 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140-M123 calibration points:
- Resistance at 50 Ω
- Resistance at 150 Ω
- Resistance at 100 Ω
Standard Certificate
HiTemp140X2-FP-72-PT-12 calibration points each channel
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140X2-FP-72-PT-52 calibration points each channel
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140X2-TD-FP-722 calibration points each channel
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140X2-TD-PT-12 calibration points each channel
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp140X2-TD-PT-52 calibration points each channel
- Temperature at 30 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
HiTemp150
2 calibration points:
- Temperature at 50 °C
- Temperature at 140 °C
ISO/IEC 17025 Accredited Certificate
IRTC101A2 calibration points:
- Temperature at 25 °C
- Temperature at 60.5 °C
1 calibration point:
- Thermocouple temperature at 0 mV(No data included on certificate)
Standard Certificate
Level1000 Channel 1:
1 calibration point:
- Temperature at 25 °C
Channel 2:
2 calibration points:
- Water level at 405 inH2O
- Water level at 775 inH2O
Standard Certificate
LF101ADevice has a digital input signal and cannot be calibrated. A declaration of conformance can be provided.Declaration of Conformance
LNDS System1 calibration point:
-Temperature 25 °C for the internal channel
3 calibration points:
0 mV for the Thermocouple channel -196°C, -80°C, 0°C
Standard Certificate
Lyotemp2 calibration points:
- Temperature at -60 °C
- Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
MicroRHTemp 1 calibration point:
- Temperature at 25 °C
Channel 2:
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25.5 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
MicroTemp1 calibration point:
- Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
Motion110Device has a digital input signal and cannot be calibrated. A declaration of conformance can be provided.
Declaration of Conformance
Motor101A (all versions)Device has a digital input signal and cannot be calibrated. A declaration of conformance can be provided.
Declaration of Conformance
OctProcess-100mAEach Channel 2 calibration points:
- Current at 0 mA
- Current at 100 mA
Standard Certificate
OctProcess-1mAEach Channel 2 calibration points:
- Current at 0 mA
- Current at 1 mA
Standard Certificate
OctProcess-25mAEach Channel 2 calibration points:
- Current at 0 mA
- Current at 24.5 mA
Standard Certificate
OctProcessV2-30mAEach Channel 2 calibration points:
- Current at 0 mA
- Current at 25 mA
Standard Certificate
OctProcessV2-160mAEach Channel 2 calibration points:
- Current at 0 mA
- Current at 100 mA
Standard Certificate
OctProcessV2-3250mAEach Channel 2 calibration points:
- Current at 0 mA
- Current at 1000 mA
Standard Certificate
OctRTDEach Channel 3 calibration points:
- Resistance at 50 Ω
- Resistance at 150 Ω
- Resistance at 100 Ω
Standard Certificate
OctRTDTempV2Each Channel 3 calibration points:
- Resistance at 50 Ω
- Resistance at 150 Ω
- Resistance at 100 Ω
Standard Certificate
OctTCTemp2000V2Each Channel 2 calibration points:
- Temperature at 25 °C
- Temperature at 60 °C
Each Channel 1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
Standard Certificate
OctTCTempV2Each Channel 2 calibration points:
- Temperature at 25 °C
- Temperature at 60 °C
Standard Certificate
OctTemp - (Serial Numbers through N1XXXX)25 °C for the internal sensor, 0 mV for the Thermocouple channels (No data included on certificate)
Standard Certificate
OctTemp (Serial Numbers beyond N1XXXX)Each Channel 2 calibration points:
- Temperature at 25 °C
- Temperature at 60 °C
- Thermocouple temperature at 0 mV (No data included on certificate)
Standard Certificate
OctTemp2000Each Channel 2 calibration points:
- Temperature at 25 °C
- Temperature at 60 °C
Each Channel 1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
Standard Certificate
OctThermoVaultEach Channel 2 calibration points:
- Temperature at 25 °C
- Temperature at 60 °C
- Thermocouple temperature at 0 mV (No data included on certificate)
Standard Certificate
OctVolt-100mV2 calibration points:
- Voltage at 0 V
- Voltage at 0.1 V
Standard Certificate
OctVolt-15V2 calibration points:
- Voltage at 0 V
- Voltage at 15.2 V
Standard Certificate
OctVolt-2.5V2 calibration points:
- Voltage at 0 V
- Voltage at 2.5 V
Standard Certificate
OctVolt-30V2 calibration points:
- Voltage at 0 V
- Voltage at 25 V
Standard Certificate
pHTemp20002 calibration points:
- Resistance at 100 Ω
- Resistance at 140 Ω
2 calibration points:
- pH voltage at 0 mV
- pH voltage at 1000 mV
Standard Certificate
PR140 (All versions)0.1 bar, 0.5 bar, 1 bar, 1.5 bar, 2 bar, 2.5 bar, 3.5 bar, 4 bar, 4.5 bar and 5.0 bar at each of the following temperature points;
20 °C, 40 °C, 60 °C, 80 °C, 100 °C, 120 °C, 140 °C Certificate only displays 1 bar, 2.5 bar and 5 bar at 20 °C and 140 °C
ISO/IEC 17025 Accredited Certificate
PR2000-1000-PSIA2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 500 psi
ISO/IEC 17025 Accredited Certificate
PR2000-100-PSIA2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 80 psi
ISO/IEC 17025 Accredited Certificate
PR2000-100-PSIG2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 80 psi
ISO/IEC 17025 Accredited Certificate
PR2000-300-PSIA2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 240 psi
ISO/IEC 17025 Accredited Certificate
PR2000-300-PSIG2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 240 psi
ISO/IEC 17025 Accredited Certificate
PR2000-30-PSIA2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 28 psi
ISO/IEC 17025 Accredited Certificate
PR2000-30-PSIG2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 28 psi
ISO/IEC 17025 Accredited Certificate
PR2000-5000-PSIA2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 500 psi
ISO/IEC 17025 Accredited Certificate
PR2000-500-PSIA2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 400 psi
ISO/IEC 17025 Accredited Certificate
PR2000-500-PSIG2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 400 psi
ISO/IEC 17025 Accredited Certificate
PRHTemp101A1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 25 % RH
- Humidity at 75 % RH
- Humidity at 50.5 % RH
2 calibration points:
- Pressure (absolute) at 0.46 bar
- Pressure (absolute) at 1.09 bar
Standard Certificate
PRHTemp20001 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 28 psi
ISO/IEC 17025 Accredited Certificate
Process101A-160mA2 calibration points:
- Current at 0 mA
- Current at 100 mA
Standard Certificate
Process101A-20mA2 calibration points:
- Current at 0 mA
- Current at 24.75 mA
Standard Certificate
Process101A-3A2 calibration points:
- Current at 0 mA
- Current at 1006 mA
Standard Certificate
PRTC110-2inH2O1 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
2 calibration points:
- Pressure (gauge) at 0.5 inH2O
- Pressure (gauge) at 1.6 inH2O
Standard Certificate
PRTC110-5inH2O1 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
2 calibration points:
- Pressure (gauge) at 1 inH2O
- Pressure (gauge) at 4 inH2O
Standard Certificate
PRTC210-1000-PSIA1 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 500 psi
Standard Certificate
PRTC210-100-PSIA1 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 80 psi
Standard Certificate
PRTC210-100-PSIG1 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 80 psi
Standard Certificate
PRTC210-300-PSIA1 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 240 psi
Standard Certificate
PRTC210-300-PSIG1 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV(No data included on certificate)
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 240 psi
Standard Certificate
PRTC210-30-PSIA1 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 28 psi
Standard Certificate
PRTC210-30-PSIG1 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 28 psi
Standard Certificate
PRTC210-5000-PSIA1 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 500 psi
Standard Certificate
PRTC210-500-PSIA1 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 400 psi
Standard Certificate
PRTC210-500-PSIG1 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 400 psi
Standard Certificate
PRTemp1000-1000-PSIA1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 500 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000-100-PSIA1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 80 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000-100-PSIG1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 80 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000-300-PSIA1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 240 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000-300-PSIG1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 240 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000-30-PSIA1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 28 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000-30-PSIG1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 28 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000-5000-PSIA1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 500 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000-500-PSIA1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 400 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000-500-PSIG1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 400 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000D-1001 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 80 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000D-301 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 28 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000D-3001 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 240 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000D-5001 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 400 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000IS-1000-PSIA1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 500 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000IS-100-PSIG1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 80 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000IS-300-PSIA1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 240 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000IS-300-PSIG1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 240 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000IS-30-PSIA1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 28 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000IS-30-PSIG1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 28.0 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000IS-5000-PSIA1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 500 psi
ISO/IEC 17025 Accredited Certificate
PRTemp1000IS-500-PSIG1 calibration point:
- Temperature at 25 °C
2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 400 psi
ISO/IEC 17025 Accredited Certificate
PRTemp140 (All Versions)2 calibration points:
- Temperature at 30 °C
- Temperature at 140 °C
77 calibration points:
0.1 bar, 0.5 bar, 1 bar, 1.5 bar, 2 bar, 2.5 bar, 3.5 bar, 4 bar, 4.5 bar and 5.0 bar at each of the following temperature points;
20 °C, 40 °C, 60 °C, 80 °C, 100 °C, 120 °C, 140 °C Certificate only displays 1 bar, 2.5 bar and 5 bar at 20 °C and 140 °C for the pressure channel
ISO/IEC 17025 Accredited Certificate
PRTrans1000-1000-PSIA2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 500 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000-100-PSIA2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 80 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000-100-PSIG2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 80 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000-300-PSIA2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 240 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000-300-PSIA2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 240 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000-30-PSIA2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 28 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000-30-PSIG2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 28 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000-5000-PSIA2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 500 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000-500-PSIA2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 400 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000-500-PSIG2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 400 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000IS-1000-PSIA2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 500 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000IS-100-PSIA2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 80 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000IS-100-PSIG2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 80 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000IS-300-PSIA2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 240 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000IS-300-PSIG2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 240 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000IS-30-PSIA2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 28 psi"
"2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 28 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000IS-30-PSIG2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 28 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000IS-5000-PSIA2 calibration points:
- Pressure (absolute) at 14.8 psi
- Pressure (absolute) at 500 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000IS-500-PSIA2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 400 psi
ISO/IEC 17025 Accredited Certificate
PRTrans1000IS-500-PSIG2 calibration points:
- Pressure (gauge) at 0 psi
- Pressure (gauge) at 400 psi
ISO/IEC 17025 Accredited Certificate
Pulse101ADevice has a digital input signal and cannot be calibrated. A declaration of conformance can be provided.Declaration of Conformance
QuadProcess-100mAEach Channel 2 calibration points:
- Current at 0 mA
- Current at 100 mA
Standard Certificate
QuadProcess-1mAEach Channel 2 calibration points:
- Current at 0 mA
- Current at 1 mA
Standard Certificate
QuadProcess-25mAEach Channel 2 calibration points:
- Current at 0 mA
- Current at 24.75 mA
Standard Certificate
QuadRTDEach Channel 3 calibration points:
- Resistance at 50 Ω
- Resistance at 150 Ω
- Resistance at 100 Ω
Standard Certificate
QuadRTDTempV2Each Channel 3 calibration points:
- Resistance at 50 Ω
- Resistance at 150 Ω
- Resistance at 100 Ω
Standard Certificate
QuadTCTemp2000V2Each Channel 2 calibration points:
- Temperature at 25 °C
- Temperature at 60 °C
Standard Certificate
QuadTCTempV2Each Channel 2 calibration points:
- Temperature at 25 °C
- Temperature at 60 °C
Standard Certificate
QuadTempEach Channel 2 calibration points:
- Temperature at 25 °C
- Temperature at 60 °C
Standard Certificate
QuadTemp2000Each Channel 2 calibration points:
- Temperature at 25 °C
- Temperature at 60 °C
Standard Certificate
QuadThermoVaultEach Channel 2 calibration points:
- Temperature at 25 °C
- Temperature at 60 °C
Standard Certificate
QuadVolt-100mVEach Channel 2 calibration points:
- Voltage at 0 V
- Voltage at 0.1 V
Standard Certificate
QuadVolt-15VEach Channel 2 calibration points:
- Voltage at 0 V
- Voltage at 15.2 V
Standard Certificate
QuadVolt-2.5V
Each Channel 2 calibration points:
- Voltage at 0 V
- Voltage at 2.5 V
Standard Certificate
QuadVolt-30VEach Channel 2 calibration points:
- Voltage at 0 V
- Voltage at 25 V
Standard Certificate
Rain101ADevice has a digital input signal and cannot be calibrated. A declaration of conformance can be provided.Declaration of Conformance
RFC1000(all versions)This device is not calibrated, declaration of conformance is availableDeclaration of Conformance
RFCO2RHTemp2000A1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
1 calibration point:
- CO2 at 400 ppm (No data included on certificate)
Standard Certificate
RFCurrent2000A-160mA2 calibration points:
- Current at 0 mA
- Current at 100 mA
Standard Certificate
RFCurrent2000A-20mA
2 calibration points:
- Current at 0 mA
- Current at 24.75 mA
Standard Certificate
RFCurrent2000A-3A2 calibration points:
- Current at 0 mA
- Current at 1000 mA
Standard Certificate
RFOT2 calibration points:
- Temperature at 15 °C
- Temperature at 85 °C
ISO/IEC 17025 Accredited Certificate
RFPRHTemp2000A1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 25.5 % RH
- Humidity at 75 % RH
- Humidity at 50 % RH
2 calibration points:
- Pressure (absolute) at 0.46 bar
- Pressure (absolute) at 1.09 bar
Standard Certificate
RFPulse2000AThis device is not calibrated, declaration of conformance is availableDeclaration of Conformance
RFRHTemp2000A1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 25.5 % RH, Temperature at 30 °C
- Humidity at 75.5 % RH, Temperature at 25 °C
- Humidity at 50 % RH, Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
RFRTDTemp2000A
3 calibration points:
- Resistance at 50 Ω
- Resistance at 150 Ω
- Resistance at 100 Ω
Standard Certificate
RFTCTemp2000A2 calibration points:
- Temperature at 25 °C
- Temperature at 60 °C
1 calibration point:
- Thermocouple temperature at 0 mV
Standard Certificate
RFTemp2000A
1 calibration point:
- Temperature at 25 °C
ISO17025 Accredited
RFVolt2000A 15V2 calibration points:
- Voltage at 0 V
- Voltage at 15.2 V
Standard Certificate
RFVolt2000A 160mVDC2 calibration points:
- Voltage at 0 mV
- Voltage at 100 mV
Standard Certificate
RFVolt2000A 2.5 VDC2 calibration points:
- Voltage at 0 V
- Voltage at 2.5 V
Standard Certificate
RFVolt2000A 30VDC2 calibration points:
- Voltage at 0 V
- Voltage at 25 V
Standard Certificate
RHTemp10001 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 25 % RH
- Humidity at 75.5 % RH
- Humidity at 50.5 % RH
ISO/IEC 17025 Accredited Certificate
RHTemp1000IS 1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 25.5 % RH, Temperature at 30 °C
- Humidity at 75.5 % RH, Temperature at 25 °C
- Humidity at 50.5 % RH, Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
RHTemp1000IS R11 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
RHTemp1000IS-SS1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 25 % RH
- Humidity at 75.5 % RH
- Humidity at 50.5 % RH
ISO/IEC 17025 Accredited Certificate
RHTemp1000SS1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
RHTemp101A1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
RHTemp20001 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
RTDTemp101A3 calibration points:
- Resistance at 50 Ω
- Resistance at 150 Ω
- Resistance at 100 Ω
Standard Certificate
Shock101-100Validation only
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
Shock101-100-EBValidation only
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
Shock101-250Validation only
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
Shock101-250-EBValidation only
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
Shock101-5Validation only
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
Shock101-50Validation only
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
Shock101-50-EBValidation only
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
Shock101-5-EBValidation only
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
State101ADevice has a digital input signal and cannot be calibrated. A declaration of conformance can be provided.Declaration of Conformance
SVR101Validation only
X, Y axis = 0 g and Z axis = 1 g
Standard Certificate
TC101A-MP2 calibration points:
- Temperature at 25 °C
- Temperature at 60.5 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
Standard Certificate
TC101A-ST2 calibration points:
- Temperature at 25 °C
- Temperature at 60.5 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
Standard Certificate
TC101A-TB2 calibration points:
- Temperature at 25 °C
- Temperature at 60.5 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
Standard Certificate
TCTemp10001 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
Standard Certificate
TCTemp20001 calibration point:
- Temperature at 25 °C
1 calibration point:
- Thermocouple temperature at 0 mV (No data included on certificate)
Standard Certificate
Temp10001 calibration point:
- Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
Temp1000FP2 calibration points:
-Temperature at 50 °C and 149 °C
Standard Certificate
Temp1000IS1 calibration point:
- Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
Temp1000IS R11 calibration point:
- Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
Temp1000IS-SS 1 calibration point:
- Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
Temp1000P2 calibration points:
- Temperature at 30 °C
- Temperature at 80 °C
Standard Certificate
Temp1000S
2 calibration points:
-Temperature at 50°C and 140°C
ISO/IEC 17025 Accredited Certificate
Temp101A1 calibration point:
- Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
TempRetriever1 calibration point:
- Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
TempRetrieverRH1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 75 % RH, Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
ThermAlert1 calibration point:
- Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
ThermAlert-GB2 Calibration points:
-Temperature at -10 °C and 70 °C
ISO/IEC 17025 Accredited Certificate
ThermAlert-P2 Calibration points:
-Temperature at -10 °C and 70 °C
ISO/IEC 17025 Accredited Certificate
ThermAlert-RH1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 25.5 % RH, Temperature at 30 °C
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 75 % RH, Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
ThermoVaultThis product is a system, you must calibrate components separately. See TC101AN/A
Titan S83 calibration points:
- Current at 0 mA
- Current at 20 mA
- Current at 10 mA
3 calibration points:
- Voltage at 0 mV
- Voltage at 100 mV
- Voltage at 50 mV
3 calibration points:
- Resistance at 0 Ω
- Resistance at 50000 Ω
- Resistance at 100000 Ω
3 calibration points:
- Resistance at 40 Ω
- Resistance at 230 Ω
- Resistance at 135 Ω
3 calibration points:
- Voltage at 0 V
- Voltage at 10 V
- Voltage at 5 V
Standard Calibration without certificate
TransiTemp-EC1 calibration point:
- Temperature at 23.5 °C
No Calibration
TransiTemp-EC-Multi1 calibration point:
- Temperature at 23.5 °C
No Calibration
TransiTempII1 calibration point:
- Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
TransiTempII-RH1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
ISO/IEC 17025 Accredited Certificate
TSR101-100Validation only,
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
TSR101-100-EBValidation only,
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
TSR101-250Validation only,
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
TSR101-250-EBValidation only,
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
TSR101-5Validation only,
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
TSR101-50Validation only,
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
TSR101-50-EBValidation only,
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
TSR101-5-EBValidation only,
X, Y axis = 0 g and Z axis = 1 g
Declaration of Conformance
ULT90-9This product is a system, you must calibrate components separately. See CryoTempN/A
ULT90-15This product is a system, you must calibrate components separately. See CryoTempN/A
UltraShock (all except Ultrashock 5 models)Shock validated to:
X and Y Axis = 0 g Z Axis = 1 g (No data included on certificate)
1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 28 psi
Standard certificate for Temperature, Humidity and Pressure Declaration of Conformance for Shock
UltraShock 5 Shock validated to:
X, Y and Z Axis = -1 g and +1 g (No data included on certificate)
1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 28 psi
Standard certificate for Temperature, Humidity and Pressure Declaration of Conformance for Shock
UltraShock 5 EBShock validated to:
X, Y and Z Axis = -1 g and +1 g (No data included on certificate)
1 calibration point:
- Temperature at 25 °C
3 calibration points:
- Humidity at 50 % RH, Temperature at 25 °C
- Humidity at 25 % RH, Temperature at 30 °C
- Humidity at 75 % RH, Temperature at 25 °C
2 calibration points:
- Pressure (absolute) at 14.75 psi
- Pressure (absolute) at 28 psi
Standard certificate for Temperature, Humidity and Pressure Declaration of Conformance for Shock
Volt101A-15V2 calibration points:
- Voltage at 0 V
- Voltage at 15.2 V
Standard Certificate
Volt101A-160mV2 calibration points:
- Voltage at 0 mV
- Voltage at 100 mV
Standard Certificate
Volt101A-2.5V2 calibration points:
- Voltage at 0 V
- Voltage at 2.5 V
Standard Certificate
Volt101A-30V2 calibration points:
- Voltage at 0 V
- Voltage at 25 V
Standard Certificate
VTMS2 calibration points:
-Temperature at 25 °C and 60 °C
3 calibration points:
Thermocouple channel -50°C, +8°C, +25°C
Standard Certificate
Wind101ADevice has a digital input signal and cannot be calibrated. A declaration of conformance can be provided.Declaration of Conformance