RHTemp1000IS

Compact DesignBattery Life Indicator

The RHTemp1000IS is an intrinsically safe, humidity and temperature data logger. Includes ISO17025 accredited calibration certificate.

Availability: 5-7 Business Days

$449.00

Extended Warranty

Description

The RHTemp1000IS is a compact, humidity and temperature logger that is FM approved as intrinsically safe for hazardous environments.

This device measures and records temperatures from -40 °C to +80 °C (-40 °F to +176 °F) with an accuracy of ±0.5 °C. The RHTemp1000IS can store up to 16,384 time and date stamped readings, and is able to be programmed to start up to 24 months in advance.

The RHTemp1000IS is compatible with the latest MadgeTech 4 Software making starting, stopping, and downloading simple and easy. Graphical, tabular, and summary data is provided for analysis and can be viewed as customizable engineering units.

The RHTemp1000IS has been Factory Mutual certified as intrinsically safe for Class I, Division 1, groups A, B, C and D and non-incendive for Class I, Division 2, groups A, B, C and D. 

General

Memory16,384 readings (all channels enabled)
Reading Rate 1 reading every second up to 1 reading every 24 hours
Battery Life 2 years typical at 25 °C, 15 minute reading intervals
Material316 Stainless Steel/Radel
Dimensions1.7 in x 0.97 in x 0.97 in (42 mm x 24.6 mm x 24.6 mm)
Operating Environment -40 °C to +80 °C (-40 °F to +176 °F), 0 %RH to 100 %RH non-condensing
Required Interface Package IFC400 or IFC406
ApprovalsCE, IS Rated

Temperature

Temperature Sensor Resistance Temperature Detector (RTD)
Temperature Range -40 °C to +80 °C (-40 °F to +176 °F)
Temperature Resolution 0.01 °C (0.018 °F)
Temperature Response
Time in Air
t63 - 8 minutes to 63% of change
t90 - 68 minutes to 90% of change
Temperature Calibrated Accuracy ±0.5 °C (0 °C to 55 °C)

Humidity

Humidity Sensor Capacitive Polymer
Humidity Range 0 %RH to 100 %RH (non-condensing)
Humidity Resolution 0.1 %RH
Humidity Response
Time in Air
t90 - 2 minutes to 90% of change
Humidity Calibrated Accuracy ±3 %RH maximum
Humidity Specified Accuracy Range25 %RH to 75 %RH
+20 °C to +40 °C
Hysteresis Error 1 % Typical,
3% Maximum