The HiTemp140 is a stainless steel temperature data logger built for use in harsh environments such as autoclaves. It is submersible, can withstand temperatures up to 140 °C (284 °F) and has an accuracy of +/-0.1 °C (0.18 °F).
The HiTemp140 data logger can store up to 32,700 readings, and features a rigid external probe capable of measuring extended temperatures up to 260 °C (500 °F). Custom probe lengths up to 7 in are available. The HiTemp140 records date and time stamped readings, and has non-volatile solid state memory. This allows the device to retain recorded data even if the battery becomes discharged.
The MadgeTech Software makes starting, stopping, and downloading the HiTemp140 simple and easy. Graphical, tabular and summary data is provided for analysis and data can be viewed in °C, °F, K or °R. The data can also be automatically exported to Excel® for further calculations.
Click here to view our probe selection chart
TEMPERATURE
Temperature Sensor | 100 Ω Platinum RTD |
---|---|
Temperature Resolution | 0.01 °C (0.02 °F) |
Temperature Calibrated Accuracy | ±0.1 °C/±0.18 °F (20 °C to +140 °C/68 °F to +284 °F) ±0.3 °C/±0.54 °F (-20 °C to +19.99 °C/-4 °F to +67.98 °F) ±0.4 °C/±0.72 ºF (-40 °C to -20.01 °C/-40 °F to +-4.02 °F) |
Probe Temperature Range | -200 °C to +260 °C (-328 °F to +500 °F) (Body of Logger Cannot Exceed 140 °C) |
GENERAL
Memory | 32,700 | |||||||
---|---|---|---|---|---|---|---|---|
Reading Rate | 1 reading every second up to 1 reading every 24 hours | |||||||
Baud Rate | 125,000 | |||||||
Battery Life | 1 year typical (1 minute reading rate at 25 °C/77 °F) User Replaceable Battery |
|||||||
Material | 316 Stainless Steel | |||||||
Dimensions |
|
|||||||
IP Rating | IP68 | |||||||
Submersible | Yes | |||||||
Operating Environment | -40 °C to +140 °C (-40 °F to +284 °F) 0 %RH to 100 %RH (Body of Logger Cannot Exceed 140 °C) |
|||||||
Required Interface Package | IFC400 | |||||||
Approvals | CE |
1 Item(s)
We have used these loggers since they were first released for validating autoclaves, and steam sterilisers, and they have performed perfectly would have absolutely no hesitation in recommending these to anyone need a self contained high temp data logger a big thumbs up from us! (Posted on 4/4/14)
1 Item(s)